Safety
Build-in back-up battery to display temperature and to operate audible and visual alarm system for up to 72 hours without AC power.
Visual and audible alarm system: High/low temperature, sensor error, power failure, door ajar, high ambient temperature, condenser clean, low battery, filter screen check, abnormal voltage.
Power failure protection: turn on delay of the cooling system when power failure, restart delay protection.
Keyboard lock, password protection to avoid set parameter randomly.
Special noise reduction design
Humanized Design
4 casters with stopper for easy moving and fixation.
Standard 25mm test hole for easy connecting to temperature monitoring systems.
3 adjustable shelves for easy storing.
Removable condenser dust gauze convenient for cleaning and maintenance.
Removable inner door for user to defrost and clean.
Standard USB port, USB port to read temperature and alarm conditions in recent 1 months.
Special
Patented technology, mixed refrigerant.
Save energy up to 50%.
Wide voltage design: 198V~242V available.
Build-in back-up battery: 72 hours power failure alarm.
9 visual and audible alarms.
Performance
Temperature Range: -40~-86°C
Ambient Temperature: +10~+32°C
Controller: Microprocessor
Display: Digital Display
Refrigeration System
Compressor/Brand: 1/SECOP
Refrigerant: Mixed
Cooling Type: Direct cooling
Defrost Type: Manual
Electrical Data
Power Supply(V/Hz): 220/50~60; 115/60
Rated Power(W): 668
Electrical Current(A): 3.5
Power Consumption: 11.2 kWh/24h
Power failure backup system: 72H
Dimensions
Capacity: 340L
Interior Size(W*D*H): 855*1006*1920mm
Exterior Size(W*D*H): 488*607*1140mm
Package Size(W*D*H): 935*1070*2128mm
N.t./G.t.(approx.): 175/205kg
Loading Qty(20’/40’/40’H): 12/24/24
Accessories
Caster: 4
USB port: Y
Inner Door/Shelf: 2/3
Test hole: 1/25MM
Remote alarm port: Optional
CO2 back up system: Optional
LN2 back up system: Optional
Cold storage box loading Qty: 19200
There are no reviews yet.